EZ Match User Manual
8-41
I/O Parameters
Tube/Cap Detection Output 1
Note: Output On under Output 1 Parameters must be set to Tube/Cap Detection (as
shown below) for this output to function.
Under Output 1 Parameters, expand the Tube/Cap Detection line.
Important: The Tube/Cap Detection Output 1 default setting for Output On is Presence.
Detection Mode
Tube Detection
Output On fields will show Tube Detection results.
Cap Detection
Output On fields will show Cap Detection results.
Output # Mode Output On
1 1 = Cap Detection Presence
2 1 = Cap Detection Unknown
3 1 = Cap Detection Absence
Definition: Detection Mode determines the type of operation to be evaluated by the
imager.
Serial Cmd: <K860,mode,output on>
Default: Cap Detection
Options: 0 = Tube Detection 1 = Cap Detection
Vedere la pagina 224