1-6 MS-3 Laser Scanner User’s Manual
1–Quick Start
Step 5 — Start Calibration
In the calibration routine, the scanner attempts various settings to determine the
optimum decode rate for the given conditions. In this process it might do the follow-
ing:
1. Start calibration:
By Serial Command
Send <@> from your terminal.
By the Test button
Press and hold the TEST button
until you hear two beeps and
see two amber performance
LEDs turn amber to indicate that
the calibration is in progress.
2. End calibration:
The test will end automatically
when the optimum combination
of settings has been achieved.
•Adjust laser power
Adjust motor speed
Adjust AGC gain
•Enable Autodiscrimination
(read several symbol types)
PWR
40%
60%
80%
20%
100%
ST
GR
11-000063-01
TEST
L
A
S
E
R
M
S
-
3
Test Button
Read rate
performance
LEDs
Vedere la pagina 17